Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 51: (1 year) CAL VER /5520 INSTRUMENT: Fluke 52: (1 year) CAL VER /5520 DATE: 26-May-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 15 NUMBER OF LINES: 97 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 51,52 Service Manual (PN 768234 February 1986). # # Compatibility: # 5520/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K 1.003 HEAD {DEGREES C} 1.004 JMPT 1.009 Fluke 52: (1 year) CAL VER /5520 1.005 HEAD DEGREES C: {K-TYPE} 1.006 MEM2 =51 1.007 DISP Connect the 5520A to the UUT using a 1.007 DISP type K SMP plug to SMP plug cable. 1.008 JMP 1.012 1.009 HEAD DEGREES C, INPUT 1: {K-TYPE} 1.010 MEM2 =52 1.011 DISP Connect the 5520A to the UUT Input 1 using a 1.011 DISP type K SMP plug to SMP plug cable. 1.012 DISP If the Fluke [MEM2] is not setup for a type K 1.012 DISP thermocouple, turn it off and then on again. 1.013 DISP Setup the Fluke [MEM2] to display degrees C if not 1.013 DISP already doing so. 1.014 5520 -182.0degC 0.9U _K TC 2W #! WARNING: Test Tol 0.9, Sys Tol 0.25, TUR 3.6000 (< 4.00). 2.001 5520 -89.0degC 0.8U _K TC 2W 3.001 5520 530.0degC 1.2U _K TC 2W 4.001 5520 1355.0degC 2.1U _K TC 2W 5.001 JMPT 9.002 Fluke 51: (1 year) CAL VER /5520 5.002 HEAD DEGREES C, INPUT 2: {K-TYPE} 5.003 DISP Connect the 5520A to the UUT Input 2 using a 5.003 DISP type K SMP plug to SMP plug cable. 5.004 DISP Select Input 2. 5.005 5520 -182.0degC 0.9U _K TC 2W #! WARNING: Test Tol 0.9, Sys Tol 0.25, TUR 3.6000 (< 4.00). 6.001 5520 -89.0degC 0.8U _K TC 2W 7.001 5520 530.0degC 1.2U _K TC 2W 8.001 5520 1355.0degC 2.1U _K TC 2W # This JMPT is necessary so repeat and cancel work correctly. 9.001 JMPT 9.005 Fluke 52: (1 year) CAL VER /5520 9.002 HEAD DEGREES C: {J-TYPE} 9.003 DISP Connect the 5520A to the UUT using a 9.003 DISP type J SMP plug to SMP plug cable. 9.004 JMP 9.007 9.005 HEAD DEGREES C, INPUT 1: {J-TYPE} 9.006 DISP Connect the 5520A to the UUT Input 1 using a 9.006 DISP type J SMP plug to SMP plug cable. 9.007 DISP Turn off the Fluke [MEM2] and turn it on again while 9.007 DISP pressing the HOLD button to select type J. 9.008 5520 -197.0degC 1.0U _J TC 2W 10.001 5520 258.0degC 1.1U _J TC 2W 11.001 5520 705.0degC 1.5U _J TC 2W 12.001 JMPF 15.001 Fluke 52: (1 year) CAL VER /5520 12.002 HEAD DEGREES C, INPUT 2: {J-TYPE} 12.003 DISP Connect the 5520A to the UUT Input 2 using a 12.003 DISP type J SMP plug to SMP plug cable. 12.004 DISP Select Input 2. 12.005 5520 -197.0degC 1.0U _J TC 2W 13.001 5520 258.0degC 1.1U _J TC 2W 14.001 5520 705.0degC 1.5U _J TC 2W 15.001 END #! T.U.R.s less than 4.00: 2 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.